Ingenia 3.0T

Philips Ingenia 3.0T Magnetic Resonance System,the first-ever digital broadband MR

PHILIPS MRI RF AMPLIFIER 3.0T AN8134 P/N: 452213203182

Xtend Magnet System

Magnet weight 4600 kg
Open bore diameter 70 cm
Maximum FOV 55 cm
Typical homogeneity at 45 cm DSV ≤ 1.1 ppm
HeliumSave technology (Zero boil-off) Yes
Cryogen boil-off rate under regular scanning conditions 0 l/hr

Omega HP gradients

Max. amplitude for each axis 45 mT/m
Max. slew rate for each axis 200 T/m/s

Resolution parameters

Max scan matrix 1024 (2048 optional)
Max. number of slices 1024
Highest in-plane resolution 5 µm

RF Transmit

Parallel RF transmission Yes
Number of independent RF amplifiers 2 AN8134
Output power 2 × 18 kW

dStream RF receive

Number of independent receive channels Channel independent
Location of analog-to-digital converter (ADC) Inside the coil
Signal chain from coil to reconstructor Fully digital
Signal chain from coil electronics to connector Digital
Signal chain from connector to magnet Digital
Signal chain from magnet to reconstructor Digital

Patient environment

Patient aperture 70 cm
Flare on both ends Yes
Tunnel diameter at both ends 95 cm
Maximum weight capacity 250 kg (550 lbs)
Patient transport system (optional) FlexTrak
Wireless patient physiological synchronization Yes
Various acoustic noise reduction solutions Yes

Site Planning

Total gantry installed weight ≤ 5800 kg
Minimum siting requirement 30 m²

dStream workflow

FlexCoverage Posterior coil Yes
FlexCoverage Anterior coils (optional) Yes
FlexConnect connectors Yes
FlexTrak tabletop Yes
FlexCaddy coil storage (optional) Yes
FlexTrak patient transport system (optional) Yes
FlexTrak Mammo mammography solution (optional) Yes

iPatient efficiency assistance

SmartStart Yes
SmartSelect Yes
SmartExam (optional) Yes
SmartLine (package dependent) Yes
SmartLink (package dependent) Yes