PHILIPS MRI RF AMPLIFIER 3.0T AN8134 P/N: 452213203182
Xtend Magnet System |
|
Magnet weight | 4600 kg |
Open bore diameter | 70 cm |
Maximum FOV | 55 cm |
Typical homogeneity at 45 cm DSV | ≤ 1.1 ppm |
HeliumSave technology | (Zero boil-off) Yes |
Cryogen boil-off rate under regular scanning conditions | 0 l/hr |
Omega HP gradients |
|
Max. amplitude for each axis | 45 mT/m |
Max. slew rate for each axis | 200 T/m/s |
Resolution parameters |
|
Max scan matrix | 1024 (2048 optional) |
Max. number of slices | 1024 |
Highest in-plane resolution | 5 µm |
RF Transmit |
|
Parallel RF transmission | Yes |
Number of independent RF amplifiers | 2 AN8134 |
Output power | 2 × 18 kW |
dStream RF receive |
|
Number of independent receive channels | Channel independent |
Location of analog-to-digital converter (ADC) | Inside the coil |
Signal chain from coil to reconstructor | Fully digital |
Signal chain from coil electronics to connector | Digital |
Signal chain from connector to magnet | Digital |
Signal chain from magnet to reconstructor | Digital |
Patient environment |
|
Patient aperture | 70 cm |
Flare on both ends | Yes |
Tunnel diameter at both ends | 95 cm |
Maximum weight capacity | 250 kg (550 lbs) |
Patient transport system (optional) | FlexTrak |
Wireless patient physiological synchronization | Yes |
Various acoustic noise reduction solutions | Yes |
Site Planning |
|
Total gantry installed weight | ≤ 5800 kg |
Minimum siting requirement | 30 m² |
dStream workflow |
|
FlexCoverage Posterior coil | Yes |
FlexCoverage Anterior coils (optional) | Yes |
FlexConnect connectors | Yes |
FlexTrak tabletop | Yes |
FlexCaddy coil storage (optional) | Yes |
FlexTrak patient transport system (optional) | Yes |
FlexTrak Mammo mammography solution (optional) | Yes |
iPatient efficiency assistance |
|
SmartStart | Yes |
SmartSelect | Yes |
SmartExam (optional) | Yes |
SmartLine (package dependent) | Yes |
SmartLink (package dependent) | Yes |